FDA Adverse Event Injury Summary report: N

NOVASURE IMPEDANCE CONTROLLED EA SYSTEM

MDR report key: 3800233 · Received April 30, 2014

Report

Report Number
1222780-2014-00070
Event Type
Injury
Date Received
April 30, 2014
Date of Event
December 16, 2013
Report Date
March 31, 2014
Manufacturer
HOLOGIC
Product Code
MNB
PMA / PMN Number
P010013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

LOT NUMBER OF THE DISPOSABLE DEVICE NOT PROVIDED BY THE COMPLAINANT, THEREFORE, THE MANUFACTURE DATE IS NOT KNOWN. DEVICE HISTORY RECORD (DHR) REVIEW AND STERILE RECORD REVIEW WERE UNABLE TO BE CONDUCTED FOR THE DISPOSABLE DEVICE AS THE LOT NUMBER WAS NOT PROVIDED BY THE COMPLAINANT. ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WARNINGS: USE CAUTION NOT TO PERFORATE THE UTERINE WALL WHEN SOUNDING, DILATING, OR INSERTING THE DISPOSABLE DEVICE. IF THE DISPOSABLE DEVICE IS DIFFICULT TO INSERT INTO THE CERVICAL CANAL, USE CLINICAL JUDGEMENT TO DETERMINE WHETHER OR NOT FURTHER DILATION IS REQUIRED. THE NOVASURE SYSTEM PERFORMS A CAVITY INTEGRITY ASSESSMENT (CIA) TEST TO EVALUATE THE INTEGRITY OF THE UTERINE CAVITY, AND SOUNDS AN ALARM WARNING OF A POSSIBLE PERFORATION PRIOR TO TREATMENT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A NOVASURE ENDOMETRIAL ABLATION ON (B)(6) 2013, THE PHYSICIAN RECEIVED AN UNSUCCESSFUL CAVITY INTEGRITY ASSESSMENT (CIA) TEST. A LAPAROSCOPY WAS PERFORMED AND THE PHYSICIAN VISUALIZED A "PERFORATION IN THE FUNDUS". THE PHYSICIAN "STITCHED" THE PERFORATION AND ABORTED THE PROCEDURE. THE PATIENT WAS DISCHARGED HOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259162 NOVASURE IMPEDANCE CONTROLLED EA SYSTEM UTERINE ABLATION DEVICE MNB HOLOGIC NS2000 UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention RADIO FREQUENCY CONTROLLER: SN# UNK