NOVASURE IMPEDANCE CONTROLLED EA SYSTEM
Report
- Report Number
- 1222780-2014-00070
- Event Type
- Injury
- Date Received
- April 30, 2014
- Date of Event
- December 16, 2013
- Report Date
- March 31, 2014
- Manufacturer
- HOLOGIC
- Product Code
- MNB
- PMA / PMN Number
- P010013
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
LOT NUMBER OF THE DISPOSABLE DEVICE NOT PROVIDED BY THE COMPLAINANT, THEREFORE, THE MANUFACTURE DATE IS NOT KNOWN. DEVICE HISTORY RECORD (DHR) REVIEW AND STERILE RECORD REVIEW WERE UNABLE TO BE CONDUCTED FOR THE DISPOSABLE DEVICE AS THE LOT NUMBER WAS NOT PROVIDED BY THE COMPLAINANT. ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WARNINGS: USE CAUTION NOT TO PERFORATE THE UTERINE WALL WHEN SOUNDING, DILATING, OR INSERTING THE DISPOSABLE DEVICE. IF THE DISPOSABLE DEVICE IS DIFFICULT TO INSERT INTO THE CERVICAL CANAL, USE CLINICAL JUDGEMENT TO DETERMINE WHETHER OR NOT FURTHER DILATION IS REQUIRED. THE NOVASURE SYSTEM PERFORMS A CAVITY INTEGRITY ASSESSMENT (CIA) TEST TO EVALUATE THE INTEGRITY OF THE UTERINE CAVITY, AND SOUNDS AN ALARM WARNING OF A POSSIBLE PERFORATION PRIOR TO TREATMENT. (B)(4).
IT WAS REPORTED THAT DURING A NOVASURE ENDOMETRIAL ABLATION ON (B)(6) 2013, THE PHYSICIAN RECEIVED AN UNSUCCESSFUL CAVITY INTEGRITY ASSESSMENT (CIA) TEST. A LAPAROSCOPY WAS PERFORMED AND THE PHYSICIAN VISUALIZED A "PERFORATION IN THE FUNDUS". THE PHYSICIAN "STITCHED" THE PERFORATION AND ABORTED THE PROCEDURE. THE PATIENT WAS DISCHARGED HOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 259162 | NOVASURE IMPEDANCE CONTROLLED EA SYSTEM | UTERINE ABLATION DEVICE | MNB | HOLOGIC | NS2000 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention | RADIO FREQUENCY CONTROLLER: SN# UNK |