FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK CR(R) PLUS DEFIBRILLATOR
MDR report key: 3800164
·
Received May 8, 2014
Report
- Report Number
- 3015876-2014-00512
- Event Type
- Malfunction
- Date Received
- May 8, 2014
- Date of Event
- April 9, 2014
- Report Date
- April 9, 2014
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K033275
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL ADVISED THE CUSTOMER THAT THEIR DEVICE IS NOT FIELD SERVICEABLE AND NO LONGER UNDER WARRANTY. IT WAS LATER CONFIRMED BY THE CUSTOMER THAT THE DEVICE HAS BEEN PERMANENTLY REMOVED FROM SERVICE. THE DEVICE HAS NOT BEEN EVALUATED BY PHYSIO-CONTROL. THE CAUSE OF THE REPORTED FAILURE COULD NOT BE DETERMINED.
Description of Event or Problem · 1
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE DISPLAYED ALL THREE ICONS (CHARGE-PAK, ATTENTION, AND SERVICE WRENCH). THE ILLUMINATION OF ALL THREE ICONS INDICATES THAT THE DEVICE MAY NOT BE ABLE TO PROVIDE SUFFICIENT DEFIBRILLATION THERAPY, IF NECESSARY. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.