ANALYTICAL E MODULE
Report
- Report Number
- 1823260-2014-03292
- Event Type
- Malfunction
- Date Received
- May 8, 2014
- Date of Event
- October 16, 2013
- Report Date
- June 10, 2014
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K961481
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS EVENT OCCURRED IN (B)(6). THE SPECIFIC PART NUMBER INVOLVED IN THIS EVENT WAS REQUESTED BUT NOT PROVIDED.
THE FOUR PATIENT SAMPLES WERE RETURNED FOR INVESTIGATION. ALL FOUR SAMPLES WERE FOUND TO BE REACTIVE FOR TOXG, CONFIRMING THE CUSTOMER'S OBSERVATIONS. THE CUSTOMER'S RESULTS WERE ALSO CONFIRMED USING A VALIDATED NEUTRALIZATION ASSAY. THE REAGENT WORKED TO SPECIFICATION.
THE CUSTOMER ALLEGED THEY RECEIVED QUESTIONABLE IGG ANTIBODIES TO TOXOPLASMA GONDII (TOXG) RESULTS FOR FOUR PATIENTS ON THEIR E-MODULE. THE CUSTOMER STATED THE SAMPLES WERE DRAWN APPROXIMATELY (B)(6) 2013. THE EXACT DRAW DATES WERE REQUESTED BUT NOT PROVIDED. THE CUSTOMER STATED THE RESULTS FROM THE E-MODULE WERE REPORTED OUTSIDE THE LABORATORY. THE SAMPLES WERE THEN REPEATED ON AN ABBOTT ARCHITECT ANALYZER. THE FIRST PATIENT'S INITIAL TOXG RESULT WAS 29.54 UI/ML AND WAS POSITIVE. THE REPEAT RESULT WAS 1.4 UI/ML AND IT WAS NEGATIVE. THE SECOND PATIENT'S INITIAL TOXG RESULT WAS 25.62 UI/ML AND WAS POSITIVE. THE REPEAT RESULT WAS 1.4 UI/ML AND IT WAS NEGATIVE. THE THIRD PATIENT'S INITIAL TOXG RESULT WAS 32.37 UI/ML AND WAS POSITIVE. THE REPEAT RESULT WAS 1.5 UI/ML AND IT WAS NEGATIVE. THE FOURTH PATIENT'S INITIAL TOXG RESULT WAS 43.5 UI/ML AND WAS POSITIVE. THE REPEAT RESULT WAS 1.0 UI/ML AND IT WAS NEGATIVE. THERE WERE NO ADVERSE EVENTS. THE TOXG REAGENT LOT NUMBER WAS 172037-02. THE EXPIRATION DATE WAS REQUESTED BUT NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 279428 | ANALYTICAL E MODULE | IMMUNOCHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |