INOGEN
Report
- Report Number
- 2124215-2014-04609
- Event Type
- Malfunction
- Date Received
- May 8, 2014
- Date of Event
- February 27, 2014
- Report Date
- April 21, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. ANALYSIS DID NOT IDENTIFY ANY DEVICE FUNCTIONAL CHARACTERISTICS THAT WOULD HAVE CAUSED THE OBSERVED HIGH PACING IMPEDANCE MEASUREMENTS.
FOLLOWING RETURN AND COMPLETION OF LABORATORY ANALYSIS, THIS EVENT WILL BE FURTHER UPDATED.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT FOLLOWING SUCCESSFUL LEAD PLACEMENT WITH FAVORABLE MEASUREMENTS ON THE PSA, HIGH OUT OF RANGE LEFT VENTRICULAR (LV) AND RIGHT VENTRICULAR (RV) VALUES WERE EXHIBITED, FOLLOWING CONNECTION WITH THIS DEVICE. TROUBLESHOOTING THE CONNECTION OF THE LEAD'S TERMINAL PIN AND HEADER PORT, FAILED TO RESOLVE THE OBSERVED OUT OF RANGE MEASUREMENTS. AS A RESULT, THE DEVICE WAS REMOVED FROM THE IMPLANT PROCEDURE AND ANOTHER BOSTON SCIENTIFIC DEVICE OF SAME MODEL, WAS SUCCESSFULLY IMPLANTED WITH NO LEAD CHANGE IMPLICATIONS. FOLLOWING CONNECTION, ALL SYSTEM MEASUREMENT WERE DEEMED ACCEPTABLE AND THE PROCEDURE COMPLETED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE ATTEMPTED IMPLANT DEVICE IS EXPECTED TO BE RETURNED FOR A POST-MARKET ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 279354 | INOGEN | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | G148 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | G148| 0296| 4678 |