FDA Adverse Event Injury Summary report: N

SUPERFLEX ASPHERIC

MDR report key: 3799540 · Received April 23, 2014

Report

Report Number
9611165-2014-00054
Event Type
Injury
Date Received
April 23, 2014
Report Date
April 14, 2014
Manufacturer
RAYNER INTRAOCULAR LENSES LTD.
Product Code
HQL
PMA / PMN Number
P060011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

RAYNER HAS ALLOCATED THE REFERENCE (B)(4) TO THIS CASE. SLIT LAMP PHOTOGRAPHS WERE PROVIDED TO RAYNER FOR REVIEW BY THE DISTRIBUTOR IN (B)(6). THE SLIT LAMP PHOTOGRAPHS RECEIVED WERE SENT TO A CLINICAL CONSULTANT FOR REVIEW. THE FEEDBACK RECEIVED FROM THE CLINICAL CONSULTANT IS AS FOLLOWS: "THE PICTURES CERTAINLY SHOW AT LEAST COATING ON THE SURFACE OF THE IOL. WHETHER THE SUBSTANCE OF THE IOL IS ALSO OPACIFIED OR NOT IS DIFFICULT TO TELL FROM THE PICTURES SUPPLIED. THE IOL IS DISLOCATED AND THERE IS PC RUPTURE. IOL SURFACE DEPOSITS AND SUBSTANCE OPACIFICATION OCCURS IN 'SICK EYES'. THIS PARTICULAR EYE HAS A VERY LONG HISTORY OF MULTIPLE PROBLEMS, MULTIPLE SURGERIES AND INTERVENTIONS AND THE DEPOSITS ON THE IOL ARE RELATED TO THESE. IN MY OPINION, FOLLOWING FACTORS MAY HAVE CONTRIBUTED TO A CHRONIC INFLAMED EYE LEADING TO SURFACE DEPOSITS ON IOL. DIABETES, PROLIFERATIVE DIABETIC RETINOPATHY, VITREOUS HAEMORRHAGE, INTRA-VITREAL INJECTIONS OF AVASTIN, POSTERIOR CAPSULE RUPTURE WHICH MUST HAVE HAPPENED DURING COMPLICATED SURGERY. EXPOSURE OF IOL TO VITREOUS ETC. WHEN IOL IS EXPOSED TO CHRONIC INFLAMMATION, VARIOUS SURGERIES, INJECTION OF PHARMACEUTICAL AGENTS SUCH AS AVASTIN, THE RISK OF SURFACE DEPOSITS INCREASES SIGNIFICANTLY. THIS CAN HAPPEN WITH ANY IOL". THE IOL HAS BEEN EXPLANTED AND RETURNED TO RAYNER.

Description of Event or Problem · 1

RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM THE DISTRIBUTOR IN (B)(6) OF AN EVENT THAT OCCURRED FOLLOWING IMPLANTATION OF A SUPERFLEX ASPHERIC 920H INTRAOCULAR LENS (IOL). THE EVENT DESCRIPTION PROVIDED STATES THAT THE HEALTHCARE PROFESSIONAL OBSERVED THE DEVELOPMENT OF OPACIFICATION IN THE POST-OPERATIVE PERIOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
245369 SUPERFLEX ASPHERIC HQL - INTRAOCULAR LENS HQL RAYNER INTRAOCULAR LENSES LTD. 920H NA

Patients

Seq Age Sex Outcome Treatment
1 66 YR Required Intervention SIX TIMES IN THE PERIOD 2010 - 2013.| AVASTIN HAS BEEN PRESCRIBED TO THE PATIENT