FDA Adverse Event Injury Summary report: N

FULL RING 180MM

MDR report key: 3794157 · Received May 6, 2014

Report

Report Number
1020279-2014-00300
Event Type
Injury
Date Received
May 6, 2014
Date of Event
April 25, 2014
Report Date
April 29, 2014
Manufacturer
SMITH & NEPHEW
Product Code
LYT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

.

Additional Manufacturer Narrative · 1

FROM THE ANALYSES CONDUCTED DURING THIS INVESTIGATION, IT WAS SHOWN THAT THE TAYLOR SPATIAL RING MOST LIKELY FRACTURED BY MECHANICAL OVERLOAD. A FRACTURE CAN OCCUR IF THE MECHANICAL LOAD APPLIED TO THE DEVICE EXCEEDED THE STRENGTH OF THE MATERIAL.

Description of Event or Problem · 1

REVISION SURGERY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
272367 FULL RING 180MM LYT SMITH & NEPHEW

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R