FDA Adverse Event
Injury
Summary report: N
FULL RING 180MM
MDR report key: 3794157
·
Received May 6, 2014
Report
- Report Number
- 1020279-2014-00300
- Event Type
- Injury
- Date Received
- May 6, 2014
- Date of Event
- April 25, 2014
- Report Date
- April 29, 2014
- Manufacturer
- SMITH & NEPHEW
- Product Code
- LYT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SF
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
.
Additional Manufacturer Narrative · 1
FROM THE ANALYSES CONDUCTED DURING THIS INVESTIGATION, IT WAS SHOWN THAT THE TAYLOR SPATIAL RING MOST LIKELY FRACTURED BY MECHANICAL OVERLOAD. A FRACTURE CAN OCCUR IF THE MECHANICAL LOAD APPLIED TO THE DEVICE EXCEEDED THE STRENGTH OF THE MATERIAL.
Description of Event or Problem · 1
REVISION SURGERY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 272367 | FULL RING 180MM | LYT | SMITH & NEPHEW |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |