FDA Adverse Event Summary report: N

MAR COR COMPLETE WATER TREATMENT SYSTEM

MDR report key: 3793897 · Received May 6, 2014

Report

Report Number
3019131-2014-00019
Date Received
May 6, 2014
Date of Event
April 14, 2014
Report Date
May 1, 2014
Manufacturer
MAR COR PURIFICATION
Product Code
FIP
PMA / PMN Number
K945559
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER NOTIFIED MAR COR (MCP) THAT THE CARBON BEDDING IN THEIR WATER PURIFICATION SYSTEM SMELT STRONGLY OF SULFUR. THE ODOR PERMEATED THE ENTIRE CLINIC AND PATIENT TREATMENT WAS DELAYED. AFTER INVESTIGATION, THE ROOT CAUSE WAS DETERMINED TO BE THAT THE CUSTOMER DID NOT DO THE REQUIRED DISINFECTION PRIOR TO REBEDDING THE TANKS AS PER MCP MANUAL ME 12565 (AUTOMATIC CARBON TANK). PROPER PROCEDURE FOR DISINFECTING THE MACHINE WERE NOT BEING FOLLOWED AND THE CARBON BEDDING SHOULD BE REPLACED. TO DATE, THERE WERE NO REPORTS OF PATIENT INJURY OR OTHER ADVERSE PATIENT IMPACT; HOWEVER, DIALYSIS TREATMENT WAS DELAYED. THIS COMPLAINT WILL CONTINUE TO BE MONITORED WITHIN MCP COMPLAINT SYSTEM.

Description of Event or Problem · 1

FACILITY PERSONNEL OBSERVED A STRONG SMELL OF SULFUR COMING FROM THE MCP CENTRAL WATER PLATFORM'S FILTRATION CARBON MEDIA. PATIENT TREATMENT WAS DELAYED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
270523 MAR COR COMPLETE WATER TREATMENT SYSTEM WATER PURIFICATION SYSTEM FIP MAR COR PURIFICATION

Patients

Seq Age Sex Outcome Treatment
1