FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3792666 · Received January 20, 2014

Report

Report Number
1720753-2014-00648
Event Type
Malfunction
Date Received
January 20, 2014
Date of Event
January 3, 2014
Report Date
January 20, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE POWER SUPPLY WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER COMPLAINED OF THE SYSTEM SHUTTING DOWN. THERE ARE NO REPORTS OF PATIENT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.