FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3792630 · Received January 20, 2014

Report

Report Number
1720753-2014-00663
Event Type
Malfunction
Date Received
January 20, 2014
Date of Event
January 2, 2014
Report Date
January 20, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE FFB (FLUOROSCOPIC FUNCTIONS BOARD) AND GIB (GENERAL INTERFACE BOARD) WERE EVALUATED AND RESEATED. THE PS1 POWER SUPPLY VOLTAGE WAS ADJUSTED TO OPTIMUM OPERATING VOLTAGE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS AN INTERMITTENT COMMUNICATION FAILURE ERROR MESSAGE. THIS ERROR IS LIKELY TO RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN DEPENDING ON WHEN THE LOSS OF COMMUNICATION OCCURRED. THERE IS NO REPORT OF A PATIENT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
46404 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1