FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3792593 · Received January 2, 2014

Report

Report Number
1314492-2014-00028
Event Type
Malfunction
Date Received
January 2, 2014
Date of Event
December 5, 2013
Report Date
December 5, 2013
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE IDENTIFIED "DOOR NOT FULLY LATCHED" ALARM WAS CONFIRMED AND REPRODUCED DURING EVALUATION. THE ALARM WAS CAUSED BY THE UPPER DOOR HOOK BEING OUT OF ADJUSTMENT. THE DEVICE WAS REWORKED TO CORRECT THIS CONDITION.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED SPECTRUM INFUSION PUMP, BAXTER EXPERIENCED A "DOOR NOT FULLY LATCHED" ALARM. ANY PATIENT INVOLVEMENT OR MEDICAL INTERVENTION IS UNKNOWN SINCE THIS WAS FOUND DURING EVALUATION OF THE DEVICE. NO ADD'L INFO IS AVAILABLE.