FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3792588 · Received January 20, 2014

Report

Report Number
1720753-2014-00677
Event Type
Malfunction
Date Received
January 20, 2014
Date of Event
January 13, 2014
Report Date
January 20, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. NO CONCLUSION CAN BE DRAWN AS REPAIR INFO IS NOT AVAILABLE AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY LOCKS UP. THIS CAUSED AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
46391 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1