SPECTRUM INFUSION PUMP

MDR report key: 3792475 · Received December 31, 2013

Report

Report Number
1314492-2013-23938
Event Type
Malfunction
Date Received
December 31, 2013
Date of Event
December 1, 2013
Report Date
December 4, 2013
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED AND REPRODUCED THE REPORTED SYMPTOM SYSTEM ERROR 105 CAUSED BY A DISLODGED CONNECTION ON THE INPUT/OUTPUT PRINTED CIRCUIT BOARD. THE INPUT/OUTPUT PRINTED CIRCUIT BOARD WAS REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP ALARMED SYSTEM ERROR 105. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.