FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3792427 · Received December 31, 2013

Report

Report Number
1314492-2013-23924
Event Type
Malfunction
Date Received
December 31, 2013
Date of Event
December 1, 2013
Report Date
December 4, 2013
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND NO EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED (THROUGH HISTORY LOG) THE SYSTEM ERROR 322 BUT COULD NOT REPRODUCE IT. THE SPECIFIC CAUSE WAS UNDETERMINED. THE UPPER AND LOWER AUXILIARY ASSEMBLIES ARE KNOWN CONTRIBUTING FACTORS TO THIS ALARM AND WERE REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A DEVICE IS ALARMING SYSTEM ERROR 322. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.