FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3792243 · Received January 13, 2014

Report

Report Number
1720753-2014-00414
Event Type
Malfunction
Date Received
January 13, 2014
Date of Event
December 23, 2013
Report Date
January 13, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE CCD CAMERA AND INTERCONNECT CABLE WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS A COMMUNICATION FAILURE ERROR MESSAGES. THIS ERROR IS LIKELY TO RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN DEPENDING ON WHEN THE LOSS OF COMMUNICATION OCCURRED. THERE IS NO REPORT OF A PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
33267 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS 9800

Patients

Seq Age Sex Outcome Treatment
1