PINNACLE SECTOR II CUP 56MM
Report
- Report Number
- 1818910-2014-18142
- Event Type
- Injury
- Date Received
- May 6, 2014
- Date of Event
- April 15, 2014
- Report Date
- April 15, 2014
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- LPH
- PMA / PMN Number
- PK000306
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL NARRATIVE:THE INVESTIGATION IS ONGOING. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PATIENT WAS REVISED TO ADDRESS SQUEAKING OF HIS HIP. THE CUP WAS FOUND TO BE VERTICAL, AND THE LINER WAS WORN, DEFORMED, AND HAD DISENGAGED FROM THE CUP. (B)(4). THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A PREVIOUS REVIEW OF THE DEVICE HISTORY RECORDS FOR THE BG5C41000 LOT CODE DID NOT REVEAL ANY RELATED MANUFACTURING DEVIATIONS OR ANOMALIES. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE REMAINING PRODUCT AND LOT COMBINATION. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED.DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
PATIENT WAS REVISED TO ADDRESS SQUEAKING OF HIS HIP. THE CUP WAS FOUND TO BE VERTICAL, AND THE LINER WAS WORN, DEFORMED, AND HAD DISENGAGED FROM THE CUP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 270610 | PINNACLE SECTOR II CUP 56MM | HIP ACETABULAR CUP | LPH | DEPUY ORTHOPAEDICS, INC. | BJ1FW1000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Required Intervention |