FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3792180 · Received May 6, 2014

Report

Report Number
1416980-2014-14542
Event Type
Malfunction
Date Received
May 6, 2014
Date of Event
April 6, 2014
Report Date
April 11, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. REVIEW OF THE EVENT LOG FOUND EVIDENCE OF THE REPORTED INCREASED INTRA-PERITONEAL VOLUME (IIPV) CONDITION. THE RETURNED HOMECHOICE HAS RECEIVED A FUNCTIONAL AND ELECTRICAL EVALUATION. THE HOMECHOICE PASSED ALL RETURNED INSTRUMENT TESTING EVALUATION PERFORMED. THE CAUSE WAS DETERMINED TO BE FALSE EMPTY DETECT AND USE ERROR, INAPPROPRIATE BYPASS OF THE CAUTION: NEGATIVE ULTRAFILTRATION (UF) ALARM. THE HOMECHOICE / HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE GIVES INSTRUCTIONS ON HOW TO BYPASS CAUTION: NEGATIVE UF ALARM. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 23:56:12. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 1552ML, INDICATING THE HOME PATIENT (HP) DRAINED 1552ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
270879 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 80 YR