FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3792134 · Received January 8, 2014

Report

Report Number
1720753-2014-00270
Event Type
Malfunction
Date Received
January 8, 2014
Date of Event
December 20, 2013
Report Date
January 8, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE GENERATOR DRIVER PCB ASSEMBLY AND FILAMENT REGULATOR WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM EXPERIENCED ERRORS, AN IMMEDIATE LOSS OF SYSTEM FUNCTIONALITY AND UN-COMMANDED SHUT DOWN. THERE IS NO REPORT OF A PATIENT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.