8800

MDR report key: 3792098 · Received January 8, 2014

Report

Report Number
1720753-2014-00295
Event Type
Malfunction
Date Received
January 8, 2014
Date of Event
December 17, 2014
Report Date
January 8, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE GIB PCB ASSEMBLY WAS EVALUATED AND REPLACED. THE SYSTEM SOFTWARE AND CALIBRATIONS WERE ALSO RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY FAILED TO ALLOW FLUOROSCOPIC EXPOSURES AND EXHIBITED INTERMITTENT FUNCTIONALITY. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12664 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 8800

Patients

Seq Age Sex Outcome Treatment
1