FDA Adverse Event
Malfunction
Summary report: N
2800
MDR report key: 3792095
·
Received January 8, 2014
Report
- Report Number
- 1720753-2014-00299
- Event Type
- Malfunction
- Date Received
- January 8, 2014
- Date of Event
- December 18, 2013
- Report Date
- January 8, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE ROTOR MOTOR PCB WAS EVALUATED AND IDENTIFIED AS REQUIRING REPLACEMENT. THE CUSTOMER DECLINED TO HAVE FURTHER SERVICE PERFORMED BY GE. NO CONCLUSION CAN BE DRAWN AS CONCLUSIVE REPAIR INFO IS UNAVAILABLE AT THIS TIME AND NO ADD'L SERVICE INFO WAS PROVIDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM FAILED OT ALLOW FLUOROSCOPIC EXPOSURES AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 12663 | 2800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 2800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |