FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3792004 · Received January 14, 2014

Report

Report Number
1720753-2014-00471
Event Type
Malfunction
Date Received
January 14, 2014
Date of Event
December 27, 2013
Report Date
January 14, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE LEMO RECEPTACLE WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO TAKE FLUOROSCOPIC IMAGE. THIS ISSUE WILL CAUSE THE SYSTEM TO BE UNUSABLE DUE TO THE INABILITY TO SEE THE LIVE IMAGE. NO PT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.