FDA Adverse Event
Malfunction
Summary report: N
7900
MDR report key: 3791870
·
Received January 6, 2014
Report
- Report Number
- 9680959-2014-00012
- Event Type
- Malfunction
- Date Received
- January 6, 2014
- Date of Event
- December 10, 2013
- Report Date
- January 6, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS GMBH
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE FOOT SWITCH WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM PRODUCED UNCOMMANDED X-RAYS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 8397 | 7900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS GMBH | 7900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |