FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3791789
·
Received January 6, 2014
Report
- Report Number
- 1720753-2014-00148
- Event Type
- Malfunction
- Date Received
- January 6, 2014
- Date of Event
- December 19, 2013
- Report Date
- January 6, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE VOLTAGE ON THE POWER SUPPLY PS2 WAS ADJUSTED TO 5.2VDC. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THERE WAS A COMMUNICATION ERROR MESSAGE. THIS ERROR MAY RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 8657 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |