FDA Adverse Event Malfunction Summary report: N

STENOSCOP

MDR report key: 3791761 · Received January 21, 2014

Report

Report Number
1720753-2014-00686
Event Type
Malfunction
Date Received
January 21, 2014
Date of Event
January 2, 2014
Report Date
January 21, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE INTERCONNECT CABLE CONNECTOR WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO PRODUCE FLUOROSCOPY X-RAY. THERE ARE NO REPORTS OF PATIENT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
51258 STENOSCOP FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) STENOSCOP

Patients

Seq Age Sex Outcome Treatment
1