FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3791736 · Received January 21, 2014

Report

Report Number
1720753-2014-00706
Event Type
Malfunction
Date Received
January 21, 2014
Date of Event
January 14, 2014
Report Date
January 21, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE SYSTEM BOARDS, CABLES, POWER SUPPLIES, FUSES, AND CONNECTORS WERE EVALUATED AND RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM EXPERIENCED AN IMMEDIATE LOSS OF SYSTEM FUNCTIONALITY AND AN UN-COMMANDED SHUT DOWN. THERE IS NO REPORT OF A PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.