FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 3791725 · Received January 21, 2014

Report

Report Number
1720753-2014-00733
Event Type
Malfunction
Date Received
January 21, 2014
Date of Event
January 9, 2014
Report Date
January 21, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND DETERMINED THAT THE SRAM CARD IN THE C-ARM WILL REQUIRE REPLACEMENT, BUT NO CONCLUSION CAN BE DRAWN AS FURTHER REPAIR INFO IS NOT AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
50993 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1