FDA Adverse Event Malfunction Summary report: N

6600

MDR report key: 3791703 · Received January 21, 2014

Report

Report Number
1720753-2014-00737
Event Type
Malfunction
Date Received
January 21, 2014
Date of Event
January 14, 2014
Report Date
January 21, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND DETERMINED THAT THE FIRMWARE ON TH MOTHERBOARD IS CORRUPTED, BUT NO CONCLUSION CAN BE DRAWN AS FURTHER REPAIR INFO IS NOT AVAILABLE.

Description of Event or Problem · 1

THE FSE REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
51024 6600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6600

Patients

Seq Age Sex Outcome Treatment
1