HOMECHOICE PRO

MDR report key: 3791571 · Received May 5, 2014

Report

Report Number
1416980-2014-14536
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
December 7, 2013
Report Date
April 11, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. REVIEW OF THE EVENT LOG FOUND EVIDENCE OF THE REPORTED IIPV CONDITION. THE CAUSE WAS DETERMINED TO BE FALSE EMPTY DETECT AND USE ERROR, CLINICIAN INAPPROPRIATELY SET MINIMUM DRAIN VOLUME PERCENT SETTING TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER'S GUIDE PROVIDES A WARNING "IF YOU SET THE MINIMUM DRAIN VOLUME % TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR THE PATIENT. THIS COULD RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 18:41:57. DURING NIGHT DRAIN CYCLE SIX, THE PATIENT'S ULTRAFILTRATION READING WAS 1553ML, INDICATING THE HOME PATIENT (HP) DRAINED 1553ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
267549 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1