FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 3791534 · Received May 5, 2014

Report

Report Number
2531779-2014-12583
Event Type
Malfunction
Date Received
May 5, 2014
Report Date
April 25, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 05/27/2014, DEVICE EVALUATION: A RETAINED SAMPLE OF THE DEVICE HAS BEEN EVALUATED BY PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: A RETAIN SAMPLE FROM THE SAME LOT NUMBER WAS TESTED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST, FORCE TEST AND FILL TEST WERE PERFORMED WITH NO FAILURES OBSERVED OR FLUID OBSERVED LEAKING OUT AT THE PLUNGER END OF THE CARTRIDGE. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. IT WAS REPORTED THAT MULTIPLE LOSS OF PRIME ALARMS HAD BEEN EMITTED BY THE PUMP.THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
266533 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION D200043

Patients

Seq Age Sex Outcome Treatment
1 19 YR