FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3791413 · Received May 5, 2014

Report

Report Number
2531779-2014-12502
Event Type
Malfunction
Date Received
May 5, 2014
Report Date
April 26, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4)-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. THE KEYPAD COVER WAS UNDAMAGED. THE PUMP POWERED UP TO THE VERIFY SCREEN WITH THE AUDIBLE BUT NO VIBRATORY FUNCTIONS. THE ¿OK¿ BUTTON RESPONDED INTERMITTENTLY WHILE THE REMAINING BUTTONS RESPONDED PROPERLY. REMOVAL OF KEYPAD COVER DID NOT FIND ANY ANOMALIES WITH THE BUTTON CONTACTS. UNRELATED TO THE BUTTON ISSUE, MOISTURE WAS OBSERVED BEHIND THE DISPLAY SCREEN AND ADDITIONAL EVIDENCE OF MOISTURE INGRESS WAS FOUND INSIDE THE PUMP WHEN IT WAS OPENED TO INVESTIGATE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT UP, DOWN, OKAY, AND CONTRAST BUTTONS WERE UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
266317 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 41 YR