FDA Adverse Event Malfunction Summary report: N

RESERVOIR 3ML

MDR report key: 3791283 · Received May 5, 2014

Report

Report Number
3004209178-2014-84273
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
April 21, 2014
Report Date
April 22, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K032005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Additional Manufacturer Narrative · 1

INSPECTED ONE OPENED AND USED RESERVOIR. PERFORMED LEAK TEST AND RESERVOIR PASSED PER SPECIFICATIONS. NO LEAKAGE ANOMALY NOTED DURING THE ANALYSIS. CHECKED O-RINGS AND STOPPER GROOVE FOR DEFECTS AND NONE WERE FOUND. RESERVOIR CONNECTED AND LOCKED INTO PLACE PROPERLY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INSULIN FROM THE RESERVOIR LEAKED INTO THE RESERVOIR CHAMBER. THE BLOOD GLUCOSE READING WAS 300MG/DL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268395 RESERVOIR 3ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-332A H7514698

Patients

Seq Age Sex Outcome Treatment
1 20 YR