FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3791281 · Received May 5, 2014

Report

Report Number
3004209178-2014-84298
Event Type
Injury
Date Received
May 5, 2014
Date of Event
April 7, 2014
Report Date
April 21, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED WITH HIGH BLOOD GLUCOSE OF 1200MG/D. TROUBLESHOOTING WAS PERFORMED, AND NO DAMAGE TO THE DRIVE SUPPORT CAP NOTED. ASSISTED THE CALLER TO RUN A MANUAL PRIME TEST AND THE INSULIN DID EXIT. THE TIME, DATE, BASAL RATES, AND BOLUS WIZARD SETTINGS WERE CORRECT. PERFORMED THE HIGH PRESSURE TEST AND PASSED. INSTRUCTED THE CALLER TO REMOVE THE CANNULA AND IT WAS BENT. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268785 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAB

Patients

Seq Age Sex Outcome Treatment
1 52 YR Hospitalization