FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3791231 · Received May 5, 2014

Report

Report Number
3004209178-2014-84345
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
April 21, 2014
Report Date
April 21, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

UNABLE TO PRIME DURING THE PRIME TEST AND INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. CRACKS TO THE DISPLAY WINDOW, BATTERY TUBE THREADS, RESERVOIR TUBE LIP AND A SCRATCHED LCD WINDOW ALSO NOTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED DURING THE PRIME PROCESS. THE BLOOD GLUCOSE READING IS 121 MG/DL. THE DRIVE SUPPORT CAP IS FLUSH. ADVISED CUSTOMER THAT THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268517 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 67 YR