FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 3791206 · Received May 5, 2014

Report

Report Number
2938836-2014-10540
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
March 19, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 41.0-41.2CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE NOTED AT 11.1-14.8CM FORM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268969 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 1581/65 NA

Patients

Seq Age Sex Outcome Treatment
1