SPL LEAD, TRANSVENOUS
MDR report key: 3791175
·
Received May 5, 2014
Report
- Report Number
- 2938836-2014-10508
- Event Type
- Injury
- Date Received
- May 5, 2014
- Date of Event
- March 26, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 22.3CM WAS RETURNED FOR ANALYSIS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.
Description of Event or Problem · 1
IT WAS REPORTED THAT DECREASED R WAVES WERE NOTED ON THE RV LEAD. DURING DEVICE CHANGE-OUT FOR NORMAL ERI, AN INSULATION ANOMALY WAS OBSERVED. THE LEAD WAS CAPPED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 268910 | SPL LEAD, TRANSVENOUS | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | SP02 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Required Intervention |