FDA Adverse Event Injury Summary report: N

RIATA ST OPTIM ACTIVE FIXATION

MDR report key: 3791169 · Received May 5, 2014

Report

Report Number
2938836-2014-10469
Event Type
Injury
Date Received
May 5, 2014
Date of Event
March 31, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BASELINE NOISE WAS OBSERVED ON THE RIGHT VENTRICULAR LEAD. THE NOISE RESULTED IN INHIBITION OF VENTRICULAR PACING. A DROP IN SENSING WAS ALSO OBSERVED. THE LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268908 RIATA ST OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7021/65 NA

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention