FDA Adverse Event Malfunction Summary report: N

UNIFY ASSURA CRT-D

MDR report key: 3791168 · Received May 5, 2014

Report

Report Number
2938836-2014-10527
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
February 20, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT NON-SUSTAINED LEAD NOISE DUE TO POST-PACED T-WAVE OVERSENSING WAS OBSERVED VIA REMOTE TRANSMISSION. PATIENT WAS ASYMPTOMATIC. THE DEVICE WAS REPROGRAMMED. NO FURTHER ISSUES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268369 UNIFY ASSURA CRT-D IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3257-40 NA

Patients

Seq Age Sex Outcome Treatment
1 60 YR