FDA Adverse Event
Malfunction
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION
MDR report key: 3791158
·
Received May 5, 2014
Report
- Report Number
- 2938836-2014-10425
- Event Type
- Malfunction
- Date Received
- May 5, 2014
- Date of Event
- March 19, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. A PARTIAL LEAD WAS RETURNED IN TWO SEGMENTS FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 15.5-16.1CM FROM THE CONNECTOR PIN. THE ETFE COATING WAS INTACT AT THIS LOCATION. (B)(4).
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 269283 | DURATA STS OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7122/60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |