DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR
MDR report key: 3791150
·
Received May 5, 2014
Report
- Report Number
- 2938836-2014-10365
- Event Type
- Injury
- Date Received
- May 5, 2014
- Date of Event
- February 22, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A ROUTINE FOLLOW UP FOR A NON-DEPENDANT PATIENT, NOISE WAS OBSERVED. LOSS OF CAPTURE AND SENSING WERE ALSO NOTED. NOISE WAS NOT REPRODUCIBLE. DIAPHRAGMATIC STIMULATION WAS OBSERVED WITH TESTING OF HIGH AMPLITUDE PACING. THE LEAD WAS EXPLANTED AND REPLACED WHEN REPOSITIONING WAS UNSUCCESSFUL. THE LEAD WAS DISCARDED IN THE FIELD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 268881 | DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7120Q/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Required Intervention |