RIATA ST ACTIVE FIXATION
Report
- Report Number
- 2938836-2014-10531
- Event Type
- Malfunction
- Date Received
- May 5, 2014
- Date of Event
- March 11, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. A LEAD WAS RETURNED IN THREE SEGMENTS FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 5.5-5.7CM FROM THE DISTAL END OF THE MIDDLE SEGMENT, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. EXTERNAL INSULATION ABRASION WAS NOTED AT 10.0-10.7CM, 10.2-10.7CM, AND 10.8-12.5CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. INTERNAL INSULATION ABRASION WAS NOTED AT 7.1-8.1CM AND 9.0-10.0CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. (B)(4).
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 268856 | RIATA ST ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7000/60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |