FDA Adverse Event Malfunction Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 3791129 · Received May 5, 2014

Report

Report Number
2938836-2014-10531
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
March 11, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. A LEAD WAS RETURNED IN THREE SEGMENTS FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 5.5-5.7CM FROM THE DISTAL END OF THE MIDDLE SEGMENT, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. EXTERNAL INSULATION ABRASION WAS NOTED AT 10.0-10.7CM, 10.2-10.7CM, AND 10.8-12.5CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. INTERNAL INSULATION ABRASION WAS NOTED AT 7.1-8.1CM AND 9.0-10.0CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. (B)(4).

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268856 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7000/60 NA

Patients

Seq Age Sex Outcome Treatment
1