FDA Adverse Event Malfunction Summary report: N

RIATA ST PASSIVE FIXATION

MDR report key: 3791120 · Received May 5, 2014

Report

Report Number
2938836-2014-10467
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
February 26, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. A PARTIAL LEAD WITH THE LEAD TIP MEASURING 42.8CM WAS RETURNED FOR ANALYSIS. INTERNAL INSULATION ABRASION WAS NOTED AT 6.8-7.0CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THIS LOCATION. (B)(4).

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268853 RIATA ST PASSIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7041/65 NA

Patients

Seq Age Sex Outcome Treatment
1