FDA Adverse Event Injury Summary report: N

FORTIFY DR, DF4 CONNECTOR

MDR report key: 3791111 · Received May 5, 2014

Report

Report Number
2938836-2014-10423
Event Type
Injury
Date Received
May 5, 2014
Date of Event
February 25, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
LWS
PMA / PMN Number
P910023
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

THE REPORTED PREMATURE BATTERY DEPLETION WAS CONFIRMED IN THE LABORATORY. BASED ON ALL AVAILABLE PARAMETER AND USAGE INFORMATION, DEVICE LONGEVITY WAS FOUND TO BE BELOW THE EXPECTED LIMITS. WITH AN EXTERNAL POWER SUPPLY ATTACHED, THE DEVICE FUNCTIONED NORMALLY. NO HIGH CURRENT WAS DETECTED DURING TESTING AND THE POWER CONSUMPTION WAS NORMAL. THE ORIGINAL BATTERY WAS RETURNED TO THE VENDOR FOR FURTHER EVALUATION AND ANALYSIS COULD NOT CONCLUSIVELY DETERMINE A ROOT CAUSE FOR THE PREMATURE BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE ER WITH COMPLAINTS OF DIZZINESS UNRELATED TO THE DEVICE. WHEN THE DEVICE WAS INTERROGATED, A SUDDEN DROP IN BATTERY VOLTAGE WITH NO RECENT SOURCES OF BATTERY DRAIN WAS NOTED. THE DEVICE WAS EXPLANTED AND REPLACED. THE PATIENT WAS DOING FINE AFTER THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268809 FORTIFY DR, DF4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST. JUDE MEDICAL, INC., CRMD CD2231-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention