FORTIFY DR, DF4 CONNECTOR
Report
- Report Number
- 2938836-2014-10423
- Event Type
- Injury
- Date Received
- May 5, 2014
- Date of Event
- February 25, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- LWS
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.
THE REPORTED PREMATURE BATTERY DEPLETION WAS CONFIRMED IN THE LABORATORY. BASED ON ALL AVAILABLE PARAMETER AND USAGE INFORMATION, DEVICE LONGEVITY WAS FOUND TO BE BELOW THE EXPECTED LIMITS. WITH AN EXTERNAL POWER SUPPLY ATTACHED, THE DEVICE FUNCTIONED NORMALLY. NO HIGH CURRENT WAS DETECTED DURING TESTING AND THE POWER CONSUMPTION WAS NORMAL. THE ORIGINAL BATTERY WAS RETURNED TO THE VENDOR FOR FURTHER EVALUATION AND ANALYSIS COULD NOT CONCLUSIVELY DETERMINE A ROOT CAUSE FOR THE PREMATURE BATTERY DEPLETION.
IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE ER WITH COMPLAINTS OF DIZZINESS UNRELATED TO THE DEVICE. WHEN THE DEVICE WAS INTERROGATED, A SUDDEN DROP IN BATTERY VOLTAGE WITH NO RECENT SOURCES OF BATTERY DRAIN WAS NOTED. THE DEVICE WAS EXPLANTED AND REPLACED. THE PATIENT WAS DOING FINE AFTER THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 268809 | FORTIFY DR, DF4 CONNECTOR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | ST. JUDE MEDICAL, INC., CRMD | CD2231-40Q | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Required Intervention |