FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA CRT-D, DF4 CONNECTOR

MDR report key: 3791073 · Received May 5, 2014

Report

Report Number
2938836-2014-10458
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
March 25, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT WAS PRESENTED WITH A NOTIFIER, AND NON-SUSTAINED LEAD NOISE DUE TO POST-PACED T-WAVE OVERSENSING WAS OBSERVED. PROGRAMMING CHANGES WERE MADE AND ISSUE WAS RESOLVED. PATIENT IS IN GOOD CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
269162 QUADRA ASSURA CRT-D, DF4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3267-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 68 YR