ATLAS PLUS HF CRT-D
Report
- Report Number
- 2938836-2014-10413
- Event Type
- Malfunction
- Date Received
- May 5, 2014
- Date of Event
- March 27, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NIK
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.
THE REPORTED FIELD EVENT OF INABILITY TO REMOVE THE ATRIAL LEAD FROM THE DEVICE WAS CONFIRMED IN THE LABORATORY. UPON RECEIPT, A PARTIAL LEAD WAS CONNECTED TO THE HEADERS ATRIAL PORT. VISUAL INSPECTION IDENTIFIED SILICONE, SEPTUM MATERIAL IN THE ARING SET SCREW HEX CAVITY. THIS MATERIAL PREVENTED FULL INSERTION OF THE TORQUE DRIVER INTO THE HEX CAVITY AND RESULTED IN DIFFICULTY LOOSENING THE SET SCREW, WHICH WAS THE CAUSE OF THE ATRIAL LEAD REMOVAL DIFFICULTY.
IT WAS REPORTED THAT DURING DEVICE CHANGE-OUT FOR NORMAL ERI, THE ATRIAL LEAD COULD NOT BE REMOVED FROM THE DEVICE. THE ATRIAL LEAD WAS CAPPED AND REPLACED DURING THE PROCEDURE. PATIENT CONDITION WAS ALL RIGHT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 268385 | ATLAS PLUS HF CRT-D | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST. JUDE MEDICAL, INC., CRMD | V-341 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |