FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3790875 · Received May 5, 2014

Report

Report Number
1416980-2014-14505
Event Type
Malfunction
Date Received
May 5, 2014
Report Date
April 10, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. A VISUAL INSPECTION WAS PERFORMED. FUNCTIONAL TESTING WAS PERFORMED WITH NO ISSUES NOTED. A REVIEW OF THE EVENT LOG WAS PERFORMED WITH THE REPORTED PROBLEM NOT BEING ABLE TO BE IDENTIFIED. A SERVICE HISTORY REVIEW SHOWED NO FAILURES/PROBLEMS THAT WERE THE SAME AS, OR SIMILAR TO, THE CURRENT DIFFICULTY. IN ADDITION, THERE WAS NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY. A DEVICE REVIEW SHOWED NOTHING RELATED TO THE MANUFACTURING PROCESS THAT WOULD CONTRIBUTE TO THE REPORTED PROBLEM. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BUT THE EVALUATION HAS YET TO BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED PROBLEM WAS VERIFIED AS A NOISY PUMP. THE PUMP WAS SCRAPPED AND THE DEVICE WAS SENT FOR SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT UNSPECIFIED PROBLEM OCCURRED ON THE HOME CHOICE (HC). THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ARRANGED THE RETURN OF THE DEVICE. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
266827 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1