FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3790854 · Received May 5, 2014

Report

Report Number
2955842-2014-02775
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
April 1, 2014
Report Date
April 9, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND A FRAYED PITCH UP/ DRIVE CABLE AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STUCK OUT AT THE INSTRUMENT'S WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE FRAYED CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A PROSTATECTOMY PROCEDURE, A BROKEN CABLE BELOW THE JAW OF THE MARYLAND BIPOLAR FORCEPS INSTRUMENT TIP WAS IDENTIFIED. NOTHING REPORTEDLY FELL INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
266810 MARYLAND BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420172-07 N10131205 184

Patients

Seq Age Sex Outcome Treatment
1