MINICAP TRANSFER SET
Report
- Report Number
- 1416980-2014-14502
- Event Type
- Malfunction
- Date Received
- May 5, 2014
- Date of Event
- April 9, 2014
- Report Date
- April 10, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- PMA / PMN Number
- K882498
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER
Narratives
(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). IN ADDITION TO A VISUAL INSPECTION, WITH THE NAKED EYE, A MICROSCOPIC INSPECTION WAS PERFORMED WITH NO ISSUES NOTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. A VISUAL INSPECTION WAS PERFORMED WITH NO ISSUES NOTED. A LEAK TEST WAS PERFORMED WITH UNDERWATER PRESSURE TESTING. A CLEAR PASSAGE TEST AND CLAMP FUNCTION TEST WERE PERFORMED WITH NO ISSUES NOTED. THE REPORTED PROBLEM WAS UNABLE TO BE DUPLICATED THROUGH A SAMPLE EVALUATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THERE WAS A LEAK COMING FROM A TWIST CLAMP OF A MINICAP TRANSFER SET WITH TWIST CLAMP. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 266340 | MINICAP TRANSFER SET | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |