FDA Adverse Event Malfunction Summary report: N

LIFEPAK CR(R) PLUS DEFIBRILLATOR

MDR report key: 3790806 · Received May 5, 2014

Report

Report Number
3015876-2014-00501
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
April 8, 2014
Report Date
April 8, 2014
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K033275
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EXAMINED THE CUSTOMER'S DEVICE AND VERIFIED THE REPORTED ISSUE. PHYSIO DETERMINED THAT THE CAUSE OF THE REPORTED ISSUE WAS DUE TO ELECTRICALLY LEAKY FILTERS, DESIGNATOR FL9 AND FL10, FROM THE ANALOG PCB ASSEMBLY. THE LEAKY FILTERS LED TO THE DEPLETION OF THE INTERNAL HYBRID LAYER CAPACITOR (HLC) BATTERIES OF THE DEVICE WHICH PREVENTED IT FROM POWERING ON.THE CUSTOMER WAS PROVIDED WITH A REPLACEMENT DEVICE.

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED FAILURE AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE DISPLAYED ALL THREE ICONS (CHARGE-PAK, ATTENTION, AND SERVICE WRENCH). THE DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
266239 LIFEPAK CR(R) PLUS DEFIBRILLATOR DEFIBRILLATORS, AUTOMATIC, EXTERNAL MKJ PHYSIO-CONTROL, INC CRPLUS

Patients

Seq Age Sex Outcome Treatment
1