ENSEAL G2 CURVED JAW
Report
- Report Number
- 3005075853-2014-03046
- Event Type
- Malfunction
- Date Received
- May 5, 2014
- Date of Event
- February 21, 2014
- Report Date
- April 28, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K112033
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). THE DEVICE WAS RECEIVED IN GOOD PHYSICAL CONDITION. UPON VISUAL INSPECTION UNDER MAGNIFICATION NO ISSUES WERE NOTED WITH THE DEVICE KNIFE. THE TIP WAS NOT BROKEN AS REPORTED BY THE CUSTOMER. THE DEVICE WAS TESTED ON THE GENERATOR AND PASSED ALL FUNCTIONAL TESTING. THE ENERGY OUTPUT DELIVERED FROM THE DEVICE WAS VERIFIED AND THE CUTTING OF THE TEST MEDIA PERFORMED AS EXPECTED. ALL THREE TONES WERE HEARD DURING FUNCTIONAL TESTING (TONE 1 IS HEARD WHEN THE ENERGY ACTIVATION BUTTON IS PRESSED; TONE 2 IS HEARD WHEN TISSUE IMPEDANCE THRESHOLD IS REACHED AND TONE 3 IS HEARD WHEN THE CYCLE IS COMPLETE). THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. THIS REPORT IS NOT INTENDED TO DENY THAT YOU EXPERIENCED A PROBLEM WITH THE DEVICE. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC TOTAL HYSTERECTOMY PROCEDURE, THE SURGEON STATED THAT THE TIP OF THE DEVICE BROKE. NO OTHER DETAILS ARE AVAILABLE. THERE WERE NO PATIENT CONSEQUENCES. CASE COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE. ONE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 267562 | ENSEAL G2 CURVED JAW | ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES | GEI | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR |