8CM HEAVY DUTY MEDIUM ATTACH
Report
- Report Number
- 1045834-2014-11958
- Event Type
- Malfunction
- Date Received
- May 5, 2014
- Date of Event
- April 18, 2014
- Report Date
- April 18, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. THE DEVICE WAS TESTED AND FOUND THAT THE TEMPERATURE WAS ABOVE SPECIFICATION. THEREFORE, THE REPORTED CONDITION WAS DUPLICATED AND CONFIRMED. THE BEARINGS WERE DAMAGED WHICH CAUSED THE HIGH TEMPERATURE. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE NORMAL WEAR OF COMPONENT PARTS OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT PRIOR TO SURGERY IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE WAS HOT. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE AS A SPARE IDENTICAL DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 268548 | 8CM HEAVY DUTY MEDIUM ATTACH | MOTOR, DRILL, ELECTRIC - ATTACHMENT | HBC | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |