FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3790608 · Received May 5, 2014

Report

Report Number
2531779-2014-12489
Event Type
Malfunction
Date Received
May 5, 2014
Report Date
April 24, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 04/24/2014 WITH THE FOLLOWING FINDINGS:THE PUMP POWERED ON TO A DIM, DISCOLORED DISPLAY SCREEN. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE DISPLAY WAS DIM AND DISCOLORED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 04/24/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268952 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 56 YR